User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Omega 2170A:Type "T" Thermocouple Thermometer DATE: 01-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 8 NUMBER OF LINES: 77 STANDARD: Omega TRC III ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ X 1.003 HEAD * ADDITIONAL REQUIRED EQUIPMENT * 1.004 DISP Omega TRC III Ice Point Cell 1.004 DISP Omega Type "T" Reference Thermocouple 1.005 HEAD INITIAL CONDITIONS 1.006 DISP Insert the reference thermocouple into the center 1.006 DISP well of the Ice Point Cell. 1.007 DISP Connect the copper end (white connector) to the 1.007 DISP 5700 OUTPUT HI/LO. Red wire is HI, green wire is LO. 1.007 DISP Connect a "T" type connector with "T" type T/C wire 1.007 DISP to UUT Channel 1 input. 1.007 DISP 1.007 DISP Connect the "T" type connector to the output of the 1.007 DISP reference thermocouple. 1.008 DISP Select Channel 1 on the MDP-8 SCANNER. 1.009 HEAD {} 1.010 HEAD {* ThermoCouple Tests *} 1.011 STD Omega TRC III 1.012 MEM2 = 1 1.013 CALL Omega 2170A Sub 1 1.014 JMP 2.001 1.015 EVAL 2.001 MEM2 = 2 2.002 CALL Omega 2170A Sub 2 2.003 CALL Omega 2170A Sub 1 2.004 JMP 3.001 2.005 EVAL 3.001 MEM2 = 3 3.002 CALL Omega 2170A Sub 2 3.003 CALL Omega 2170A Sub 1 3.004 JMP 4.001 3.005 EVAL 4.001 MEM2 = 4 4.002 CALL Omega 2170A Sub 2 4.003 CALL Omega 2170A Sub 1 4.004 OPBR Is the UUT DTB# 12-187? 4.005 JMPT 8.005 4.006 JMP 5.001 4.007 EVAL 5.001 MEM2 = 5 5.002 CALL Omega 2170A Sub 2 5.003 CALL Omega 2170A Sub 1 5.004 JMP 6.001 5.005 EVAL 6.001 MEM2 = 6 6.002 CALL Omega 2170A Sub 2 6.003 CALL Omega 2170A Sub 1 6.004 JMP 7.001 6.005 EVAL 7.001 MEM2 = 7 7.002 CALL Omega 2170A Sub 2 7.003 CALL Omega 2170A Sub 1 7.004 JMP 8.001 7.005 EVAL 8.001 MEM2 = 8 8.002 CALL Omega 2170A Sub 2 8.003 CALL Omega 2170A Sub 1 8.004 HEAD {} 8.005 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.